loader
  • Input from Rtd (Pt100, Pt 1000, Ni100), TC (B,E, J, K, N, R, S, T) and mV (-10… +70 mV);
  • potentiometer (0… 400 W/0… 4 kW);
  • three custom linearizations (up to 96 segments);
  • Reversible output (e.g.: 20… 4 mA);
  • Accuracy : better than 0.2% fs;
  • Configurable by PC plus RF programmer or by smartphone (using a free of charge app);
  • Cold junction compensation disabling;
  • Programmable retransmitted range;
  • Operating temperature: -40… +85°C;
  • Compliance with CE, EN 61326-1;
  • Calibration according to EN 60584-1: 2013, EN60751: 2008.
  • Output Signal : 4-20mA
  • Max Load : max. (V power supply-7.5V)/0.022A (current output)
  • Measurement : tempt-linearity,resistance
  • Circuit limit : <=22Ma
  • Response time : <=1second
  • Saturation current : Low side 3.9mA, high side 20.5Ma
  • Alarm current : sensor damger or sensor turnoff output is 3.9mA or 22mA(except TC)
  • Accuracy : 0.1%FS
  • Measuring accuracy is related to measuring range
  • Power supply : U=12V to 40V
  • Working temperature : -40 to 85°C
  • Storage temperature : -40 to 100°C
  • Condenstaion : Allowable
  • Protection : IP00; IP66 (mounted)
  • Earthquake resistance : 4g/2 to 150HZ
  • Voltage effection : can ignore
  • Mounted angle : no limit
  • Mount area : top of B type box
  • Model no. MI-213, ATT1

Leave a Reply

Your email address will not be published. Required fields are marked *

Inquire Now